Tuesday, February 16, 2021

Pre‐clinical evaluation of a novel robotic system for transoral robotic surgery.

xlomafota13 shared this article with you from Inoreader

Abstract

The Versius Surgical System (CMR Surgical, Cambridge, UK) is a new robotic surgical platform that is CE marked and available for clinical use.

This is the first pre‐clinical evaluation of the Versius Surgical System in accordance with phase 0 of the IDEAL‐D Framework.

Optimal setup of the Versius Surgical System for use in transoral surgery was evaluated and defined.

Two experienced TORS surgeons performed 3 index procedures – lateral oropharyngectomy, tongue base mucosectomy and partial supraglottic laryngectomy using three fresh frozen cadavers. Fourteen resections were successfully completed establishing feasibility.

Instrument tremor occurred due to suboptimal distal instrument stabilisation. A bespoke transoral stabilisation port is proposed to address this issue. Further preclinical assessment is required prior to phase 1 IDEAL‐D trial evaluation.